COMPANY PRODUCTS TECHNOLOGY COMMUNITY SERVICES
Home > Á¦Ç°Á¤º¸ > RF Shielded Anechoic Chamber  
SpaceSaver¢â 26H
Key Features:
Excellent Site Attenuation Correlation for Emissions Pre-Scans
26 MHz to 18 GHz, RI Measurements
30 MHz to 18 GHz, RE Pre-Scans
Radiated Immunity Applications: IEC 61000-4-3, EN 61000-4-3 RI, ANSI C63.4-1992 (Pre-Scan)
High Field Uniformity over Operating Frequency (Immunity Testing)
Turnkey Chamber Solutions
 
 
 
 
 
 
 
 
 
 
 
 
 
 
The SpaceSaverTM 26H is a space saving, multi-use EMC test chamber designed to perform IEC 61000-4-3, and EN 61000-4-3 RI compliance measurements. The SpaceSaverTM 26H can also be used for ANSI C63.4-2000 RE pre-scans (with limited scan Height.
Turnkey Systems

ETS-Lindgren is an integrated manufacturer producing such well-known brands as ETS-Rantec absorber, Series 81 and ETS-Euroshield shielding and RF doors, ETS-EMCO antennas, towers, turntables, and test accessories. All of our products are designed to work together for maximum efficiency and performance.
Standard Configuration
  • One 3.81 cm (1.5") pipe penetration with flange nuts and cap
  • Interior nominal shield-to-shield dimensions of 7.3 m x 3.0 m x 3.0 m (22' x 10' x 10') (.914   m w x 2.13 m h(3 ft x 7 ft)
  • One single-leaf, Series 201, recessed contact mechanism (RCM), manually operated,   RF-sheilded door, .91 m x 2.13 m (3' x 7') clear opening
  • One door maintenance kit
  • Dielectric vapor barrier and masonite underlayment
  • 3 mm (1/8") anti-static vinyl floor tile
  • Two waveguide vents, 30.48 cm x 30.48 cm (12' x 12') in the ceiling
  • Four 2 x 30 amp, 60 Hz power-line filters for lights and EUT (wiring not included)
  • Modular shielded room and support structure
  • Two "N" type connectors
  • Guaranteed performance and a five-year limited warranty; one-year warranty on doors,   filters and moving parts; two year warranty on optional ETS-Lindgren's EMCO brand   equipment
  • One threaded brass ground stud 1.27 cm (0.5") diameter x 12.7 cm (5") long
  • Four corner-mounted light fixtures with floodlights (electrical distribution not included)
  • Shield test per MIL-STD-285, one frequency test at 1 GHz
  • ETS-Lindgren's Rantec brand absorber treatment on chamber ceiling and four walls
  • ETS-Lindgren's Rantec brand FT-100 on Series 201 door
  • ETS-Lindgren's Rantec brand partial absorber treatment for the floor: 16 pieces each of   FT-100c and EHP-18PCL on an area of the chamber floor
  • Field uniformity calibration per IEC 61000-4-3 from 26 MHz to 1 GHz
  • Installation of the enclosure and absorber
  • Two connector panels, 15.24 cm x 30.48 cm (6" x 24") clear opening
    • Options
    • Raised reflective ground plane floor system
        - 15.24 cm (6") height, requires additional 15.24 cm (6") additional ceiling height
        - 30.38 cm (12") height, requires additional 30.48 cm (12") ceiling height
    • Site Attenuation Testing per ANSI C63.4-2000 to 1 GHz
    • Shielded control room 3.05 m x 3.05 m x 2.44 (10' x 10' x 8')
    • Fire detection and suppression systems
    • CCTV monitoring system
    • Intercom system
    • Electrical distribution
    • Additional filtering
    • Room brightening covering over FT-100
    • Various sizes of series 201 RF shielded swing doors
    • Sliding doors for flush-threshold opening
    • Structural design calculations and certification
      • Specifications
        ABSORBER
        TYPE
        INTERNAL
        WORKING
        AREA
        SHIELD
        ROOM
        INNER DIM
        OVERALL
        STRUCTURE
        DIM
        FT-100 Ferrite Tile
        FT-100c Ferrite Tile
        (Carpeted)
        FS-400 FerroSorb
        (L) 6.92 m
        .... 22.7 ft (W) 2.25 m
        .... 7.38 ft
        (H) 2.65 m
        .... 8.69 ft
        (L) 7.32 m
        24.0 ft (W) 3.05 m
        10.0 ft
        (H) 3.05 m
        10.0 f
        (L) 7.49 m
        24.6 ft (W) 3.22 m
        10.6 ft
        (H) 3.2 m
        10.5 ft
        FREQUENCY
        RANGE
        APPLICABLE
        STANDARDS
        IEC 1000-4-3
        FIELD
        UNIFORMITY
        ANSI C63.4
        NORMALIZED SITE
        ATTENUATION
        26 MHz to 18 GHz IEC 61000-4-3
        EN 61000-4-3
        ANSI C63.4 prescan
        (limited height)
        26 MHz - 18 GHz
        1.5 m x 1.5 m
        +6dB
        75% 16 points
        30-300 MHz +/- 6dB
        300-1000 MHz +/- 4dB
        1-18 GHz +/- 4dB
        1.5 m dia 1m h